Instrument 02 of 06
Seven errors add as areas.
One square meets the limit.
Each contribution to a measurement is a square. Their areas add into one combined square, and the combined side is compared with the legal limit.
Instrument 02
Seven small errors add as areas, so the largest square decides the result.
Set each contribution as a standard uncertainty in percent of reading, choose the limit and read the verdict. Defaults are illustrative; a contribution you cannot assess is unknown, never zero.
Contributions, percent of reading
IllustrativeSquares 1 to 7, one per contribution; below, the combined square (solid) against the limit (dashed), on one scale so that areas compare. A dashed contribution is unknown and left out of the sum.
This is an illustration, not advice. It certifies no meter and no installation, it replaces no verification by the legal-metrology body, and every uncertainty is an input you set.
Why squares
Areas add and sides do not, so the small contributions barely move the result.
A fixed, illustrative drawing: the seven default contributions added one at a time, largest first.
The largest contribution is one square.
The meter factor stands at 0.10 percent of reading in the default budget. Alone, it is a combined uncertainty of 0.100 and nothing else.
A second error adds its area, not its side.
Density at 0.08 does not lift the total to 0.18. Its area joins the first, and the combined side reads 0.128. Independent errors partly cancel.
Three squares already carry 74% of the area.
Proving joins them. The combined side reads 0.146. Most of the budget now sits in three contributions, and the rest sit in the margins.
The four small squares move the side by almost nothing.
Sampling, temperature, pressure and timing take it from 0.146 to 0.169. Perfecting them would be effort spent where the area is not.
The limit is a square too, and the budget sits inside it.
The whole-system limit is 0.3%0.3%Maximum permissible error of an accuracy class 0.3 dynamic liquid measuring system (custody transfer of crude oil and products).Source: OIML R 117-1:2019, Dynamic measuring systems for liquids other than water. Open the publication ↗. The headroom is the distance between the two sides, and an over-limit square holds the registration.
What to improve
Improve the largest square first: it buys more headroom than the three smallest together.
Reduce one contribution at a time and see how much of the combined uncertainty it removes. Gains are in points of reading, on your own inputs.
| Contribution | Gain | Points |
|---|---|---|
| 1. Meter factor | 0.024 | |
| 2. Temperature | 0.006 | |
| 3. Pressure | 0.002 | |
| 4. Density | 0.015 | |
| 5. Sampling | 0.008 | |
| 6. Proving | 0.011 | |
| 7. Timing | 0.001 |
Illustrative Bar length: gain against a full scale of 0.12 points.
Headroom is the gap between the combined uncertainty and the limit. Reducing a contribution helps in proportion to its area, so the squares that look small in the drawing return little for the work they cost.
The ranking depends on your inputs. Raise the density contribution and it takes the lead; mark one unassessed and its gain reads n/a, because an unknown cannot be improved.
The registration gate
An over-limit budget and an incomplete budget both hold the registration.
No record, no fee: until the budget is complete and inside the limit, an event stays provisional and earns nothing. Try the three outcomes.
Three outcomes
IllustrativeThe three buttons set the same inputs as the instrument above, so the squares there move too. A partial sum can look comfortable while a contribution is still missing, which is why an incomplete budget holds the registration whatever the number reads.
Holding is not a penalty. It keeps an unproven reading out of the public record until someone has proved it.
The working
Every step of the sum is shown, so an auditor can replay it.
The ledger recomputes with your inputs. The standards beside it are the ones the model is read against.
| Contribution | u | u squared | Share of area |
|---|---|---|---|
| 1. Meter factor | 0.10 | 0.0100 | 35% |
| 2. Temperature | 0.05 | 0.0025 | 9% |
| 3. Pressure | 0.03 | 0.0009 | 3% |
| 4. Density | 0.08 | 0.0064 | 22% |
| 5. Sampling | 0.06 | 0.0036 | 13% |
| 6. Proving | 0.07 | 0.0049 | 17% |
| 7. Timing | 0.02 | 0.0004 | 1% |
| Sum of squares | 0.0287 | ||
| Square root: combined | 0.169 |
Dynamic measuring systems for liquids other than water.
Petroleum measurement: standard reference conditions.
General requirements for the competence of testing and calibration laboratories.
American Petroleum Institute Manual of Petroleum Measurement Standards.
The reference conditions of a petroleum measurement are explicit: 15 °C and 101.325 kPa15 °C and 101.325 kPaStandard reference conditions for petroleum and liquid petroleum products.Source: ISO 5024:1999, Petroleum liquids and liquefied petroleum gases, measurement, standard reference conditions. Open the publication ↗. A volume without them is not a record.
Method and limits
Four rules fix the arithmetic, and four limits say what it cannot do.
Standard uncertainty
Each contribution is entered as a standard uncertainty in percent of reading: meter factor, temperature, pressure, density, sampling, proving and timing.
Square
Each is squared. The square is its area, in the same units as every other.
Add
The areas are summed. Independent contributions add this way, which is why they partly cancel.
Root and compare
The square root of the sum is the combined uncertainty, compared with the limit. Over the limit, or any contribution unknown, and the registration is held.
Maximum permissible error of an accuracy class 0.3 dynamic liquid measuring system (custody transfer of crude oil and products)
OIML R 117-1:2019
0.3%Maximum permissible error of an accuracy class 0.3 dynamic liquid measuring system (custody transfer of crude oil and products).Source: OIML R 117-1:2019, Dynamic measuring systems for liquids other than water. Open the publication ↗Maximum permissible error of the meter alone in the same accuracy class
OIML R 117-1:2019
0.2%Maximum permissible error of the meter alone in the same accuracy class.Source: OIML R 117-1:2019, Dynamic measuring systems for liquids other than water. Open the publication ↗Standard reference conditions for petroleum and liquid petroleum products
ISO 5024:1999
15 °C and 101.325 kPaStandard reference conditions for petroleum and liquid petroleum products.Source: ISO 5024:1999, Petroleum liquids and liquefied petroleum gases, measurement, standard reference conditions. Open the publication ↗Independence is assumed
Root sum of squares is exact only for independent contributions. Correlated errors need a covariance term that this model does not carry.
No coverage factor
The combined standard uncertainty is compared with the limit as it stands. Where the rules call for an expanded uncertainty, that is a further step.
A class limit, not a certificate
The two limits are maximum permissible errors of a class of system. Meeting them in a model certifies no installation.
Tested arithmetic
The model has automated tests: a 3 and 4 pair combine to 5, areas add, an unassessed contribution makes the budget incomplete, and an exact tie with the limit is within it.
Where to go next